Laser Induced Fluorescence Detection
Operation
The P/ACE System MDQ Instrument Configuration screen allows the user to define the types of
trays installed, define the home position of each tray, identify the firmware version and access
the LIF Calibration Wizard.
Figure 15 Configuring the LIF Detector
5
Choose OK to accept all the parameters set up in the configuration.